- 탄성 블리스터 액츄에이터를 이용한 일회용 미세유체 주입펌프
- ㆍ 저자명
- 장웅기,김형진,김병희,서영호,Jang. Woong Ki,Kim. Hyung Jin,Kim. Byeong Hee,Seo. Young Ho
- ㆍ 간행물명
- 한국생산제조시스템학회지
- ㆍ 권/호정보
- 2013년|22권 2호|pp.235-240 (6 pages)
- ㆍ 발행정보
- 한국생산제조시스템학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
This paper presents a disposable microfluidic infusion pump using the restoring force of elastomeric membrane of Polydimethylsiloxane. Microfluidic infusion pump consists of hydraulic resistance control part, elastomeric blister actuator part, drug inlet and outlet. Expanded elastomeric blister actuator continuously pushes liquid in the chamber to outlet. At same time, microchannel diameter near the outlet was controlled by thin elastomeric membrane in hydraulic resistance control part. Eventually flow rate of infused liquid is controlled by air pressure. In experimental study, the amount of the filled liquid in the blister is precisely controlled by the height of the blister. Flow rate of infused liquid could be controlled, that is, controlled release of the drug over time was possible by adjusting hydraulic resistance and restoring pressure with the blister actuator.