- 챔버 내 발생 입자를 이용한 CPC 수농도 비교 평가
- ㆍ 저자명
- 배민석,박다정,박승식,Bae. Min-Suk,Park. Da-Jeong,Park. Seung-Shik
- ㆍ 간행물명
- 한국대기환경학회지
- ㆍ 권/호정보
- 2014년|30권 6호|pp.619-630 (12 pages)
- ㆍ 발행정보
- 한국대기환경학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
Two butanol-based Condensation Particle Counters (CPC 3022, CPC 3025), three water-based CPCs (CPC3781, CPC3785${ imes}$2), a Gardner Counter, a Fast Mobility Particle Sizer (FMPS), and an Aerosol Electrometer (AE) were deployed to measure the number concentrations from atomized aerosol under six different conditions. Comparisons of particle number concentrations measured by the CPCs, FMPS, and AE were conducted to evaluate the performance of the each CPCs using laboratory generated artificial particles such as NaCl, succinic acid ($C_4H_6O_4$), and particles generated by propane torch & heat gun in the chamber. Good correlation between the CPC3025 and FMPS was observed for the total particle number concentrations in the size range 15 nm to 90 nm. In addition, this paper suggests that photometric mode in water-based CPC3785 could not be used as quantitative of number concentrations for CPC3785.