- 방사선사진용 디지털 영상시스템의 정량적 평가에 관한 실험적 연구
- ㆍ 저자명
- 조형희,김은경,Cho. Heang-Hee,Kim. Eun-Kyung
- ㆍ 간행물명
- 齒科放射線
- ㆍ 권/호정보
- 1994년|24권 1호|pp.137-147 (11 pages)
- ㆍ 발행정보
- 대한구강악안면방사선학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
A digital imaging system using Machintosh Ⅱ ci computer, high resolution Sony XC-77 CCD camera, Quickcapture Frame Grabber Board was evaluated for quantitative analysis of standardized periapical film with aluminum step wedge. The results were as follows: 1. Correlation between Al thickness and gray level was high-positively associated(r²=0.99, p<0.001). 2. Correlation between measured weight of experimental lesion and estimated relative lesion volume by digital subtracted radiography was also high-positively associated (r²=0.98, p<0.001). 3. As exposure time was increased, mean gray level was decreased(p<0.01) and slope of regression line between Al thickness and gray level was also decreased (p<0.01). And when the exposure time was shorter than 0.2 second, the value of r² was relatively low. On the basis of the above results, it is considered that this digital imaging system using a Macintosh Ⅱ ci computer & a high resolution CCD monochrome camera will be useful in evaluating digitized image from standardized periapical film quantitatively.