  • $^{99m}Tc$용 콜리메타의 성능과 SPECT 화상의 영향
  • Comparison of SPECT Images with $^{99m}Tc$ Collimators
저자명
이만구,이정옥,박성옥,Lee. Man-Koo,Lee. Jeong-Ok,Park. Soung-Ock
간행물명
대한방사선기술학회지
권/호정보
2001년|24권 2호|pp.35-40 (6 pages)
발행정보
대한방사선과학회
파일정보
정기간행물|
PDF텍스트
주제분야
기타
이 논문은 한국과학기술정보연구원과 논문 연계를 통해 무료로 제공되는 원문입니다.
서지반출

기타언어초록

Performance of SPECT imaging systems which use a rotating gamma camera, are affected by characteristics of the detector-collimator assembly, the data acquisition method, and the filter used in imaging reconstruction. The purpose of this study Is to examine image qualifies of SPECT with different types of low energy collimators. The SPECT imaging system in this study is a digital gamma camera system GCA-901A(Toshiba) and a data processing unit Scintipac-700(Shimadzu). The four types of collimators compared are UHR(ultra high resolution), LEHR(low energy high resolution), LEGP(low energy general purpose), and I-123 PAR(Parallel), with 0.27, 0.66, 1.00, and 2.06 relative sensitivity, respectively. In this case of the same collimators, the spatial resolutions measured in the slice plane showed a slight difference in the FWHM values(mean values of UHR, LEHR, LEGP, and I-123 PAR were 11.3 mm, 13.6 mm, 15.8 mm, and 20.4 mm, respectively) between the center and the circumference of the field of view, in the radial direction, but a large difference in the tangential direction, with lower FWHM values(values of UHR, LEHR, LEGP, and I-123 PAR were 8.4 mm, 8.7 mm, 9.3 mm, and 10.8 mm at 12 cm from the center, respectively). In comparison of SPECT images with the four types of collimators, except for the I-123 PAR collimator, image qualities of UHR, LEHR, and LEGP collimators showed only a slight difference. From the point of for, it is expected that the LEGP collimator would be suitable for SPECT imaging with $^{99m}Tc$.