  • 가상쐐기와 기존쐐기의 물리적 특성 비교
  • Physical Characteristics Comparison of Virtual Wedge Device with Physical Wedge
저자명
조정근,최계숙,임청환,김정구,정홍량,이정옥,이만구,Cho. Jung-Keun,Choi. Kye-Sook,Lim. Cheong-Hwan,Kim. Jeong-Koo,Jung. Hong-Ryang,Lee. Jung-Ok,Lee. Man
간행물명
대한방사선기술학회지
권/호정보
2001년|24권 2호|pp.49-52 (4 pages)
발행정보
대한방사선과학회
파일정보
정기간행물|
PDF텍스트
주제분야
기타
이 논문은 한국과학기술정보연구원과 논문 연계를 통해 무료로 제공되는 원문입니다.
서지반출

기타언어초록

We compared the characteristics of Siemens virtual wedge device with physical wedges for clinical application. We investigated the characteristics of virtual and physical wedges for various wedge angles (15, 30, 45, and 60) using 6- and 15-MV photon beams. Wedge factors were measured in water using an ion chamber for various field sizes and depths. In case of virtual wedge device, as upper jaw moves during irradiation, wedge angles were estimated by accumulated doses. These measurements were performed at off-axis points perpendicular to the beam central axis in water for a $15;cm;{ imes};20;cm$ radiation field size at the depth of 10 cm. Surface doses without and with virtual or physical wedges were measured using a parallel plate ion chamber at surface. Field size was $15;cm;{ imes};20;cm$ and a polystyrene phantom was used. For various field sizes, virtual and physical wedge factors were changed by maximum 2.1% and 3.9%, respectively. For various depths, virtual and physical wedge factors were changed by maximum 1.9% and 2.9%, respectively. No major difference was found between the virtual and physical wedge angles and the difference was within 0.5. Surface dose with physical wedge was reduced by maximum 20% (x-ray beam : 6 MV, wedge angle : 45, SSD : 80 cm) relative to one with virtual wedge or without wedge. Comparison of the characteristics of Siemens virtual wedge device with physical wedges was performed. Depth dependence of virtual wedge factor was smaller than that of physical wedge factor. Virtual and physical wedge factors were nearly independent of field sizes. The accuracy of virtual and physical wedge angles was excellent. Surface dose was found to be reduced using a physical wedge.