Sodium hypochlorite is used worldwide as a water disinfectant and in bleaching agent. Sodium hypochlorite applied to water initially undergoes hydrolysis to form free chlorine consisting of hypochlorous acid(HOCl) and hypochlorite ion($OCl^-$). For free chlorine determination, an electrochemical method is simple due to the electroactivity of free chlorine; it measures current and is free of most reagents. Amperometric free chlorine sensor has been developed with gold (Au)-based electrode. The 3-electrode free chlorine sensor whose working and counter electrodes were Pt exhibited excellent response to HClO at +400mV vs. Ag/AgCl/sat. KCl. In addition, the use of a pH error correction algorithm provided a reliable measurement of residual free chlorine in water sample without any pretreatment in the normal pH range(pH 6~8) of municipal water supply. The free chlorine sensor installed in on-line monitoring system could be used to continually monitor the level of residual free chlorine in real samples.